Replacing and Removing Electrodes:
Replace QUIK-COMBO electrodes after 50 defibrillation shocks or 24
hours on the patient’s skin or 8 hours of continuous pacing.
To remove QUIK-COMBO or FAST-PATCH electrodes from the patient: 1
Slowly peel back the electrode from the edge, supporting the skin.
Removing
Therapy Electrodes from Skin
2 Clean and dry
the patient’s skin. 3 When replacing electrodes, adjust the
electrode positions slightly to help prevent skin burns. 4 Close the
protective cover on the QUIK-COMBO therapy cable connector when the
cable is not in use. In addition, disconnect the defibrillation
cable from the FAST-PATCH electrodes. 1 Press down around the
electrode post. 2 Pinch the snap connector with the fingers of the
other hand and pull straight up.